HART and Fieldbus communicator: field configuration, calibration and diagnostics with the Additel ADT295

A HART and Fieldbus communicator is the portable instrument a technician uses to talk to installed transmitters and other process instruments, in order to configure, check and diagnose them without removing them from the plant. The Additel ADT295 is a field communicator that supports HART, FOUNDATION Fieldbus and Profibus PA, works both online and offline, and includes a current and voltage calibrator with 24 V loop power supply.

Let's see how a multiprotocol communicator works, which tasks it covers in the field and in the laboratory, and which technical features are worth considering before choosing one.

What is a HART and Fieldbus communicator and what is it used for?

A HART and Fieldbus communicator (or field communicator) is a portable instrument that connects to a process device and reads or changes its parameters through the device's digital communication protocol. In practice, it allows you to:

  • read the status and measurements of a transmitter, a positioner or another field instrument;
  • change configuration parameters (measuring range, units, damping, output);
  • run the test methods and adjustment procedures provided by the manufacturer;
  • diagnose loop, wiring or instrument faults.

The communicator is the typical working tool for process instrumentation maintenance in the oil & gas, chemical, energy and process industries, as well as in metrology laboratories and instrument workshops.

HART, FOUNDATION Fieldbus and Profibus PA: three protocols in a single instrument

Different generations of instrumentation coexist in plants. A communicator that supports several protocols avoids the need to keep a different instrument in stock for each technology.

HART protocol

The HART communicator uses the HART protocol, which superimposes a digital signal (FSK modulation) on the 4-20 mA analogue signal without disturbing it. The instrument therefore keeps transmitting the measurement while the technician reads and configures its parameters. To work in a loop, a HART communicator generally requires a loop resistance of about 250 Ω: in the ADT295 HART communicator, the internal resistor is already built in, so no external resistors are needed.

FOUNDATION Fieldbus and Profibus PA

FOUNDATION Fieldbus and Profibus PA are digital fieldbuses in which several instruments share the same cable and communicate bidirectionally. Both use a physical layer at 31.25 kbit/s, in accordance with the IEC 61158-2 standard. Being able to work with all three protocols makes it possible to cover both plants with 4-20 mA loops and those with bus-based instrumentation.

How to configure a HART transmitter: online and offline

The configuration of HART transmitters is based on DDs (Device Descriptions), the files that describe the parameters and functions of each instrument. The communicator uses them to display menus, parameters and test methods consistent with those of the manufacturer.

The ADT295 works in two modes:

  • online, communicating directly with the devices installed in the plant while they are in operation;
  • offline, in the laboratory or workshop, to prepare and check instruments before installation.

Available functions include: DD menu navigation, parameter configuration, test methods, data trends, customisable hotkeys, favourites, configuration storage, offline simulation, full parameter backup and restore, and connection history. Saving a configuration and reapplying it to several instruments of the same type reduces commissioning times and data-entry errors.

Current and voltage calibrator and loop diagnostics in a single instrument

A multifunction communicator brings together in a single device functions that previously required several instruments.

Current and voltage calibrator

The Additel ADT295 HART communicator includes a current and voltage calibrator: it measures voltage with automatic ranging, and measures and generates current with an accuracy of 0.01% of reading (plus 1.5 µA on current). It is used to simulate an input signal to a receiver, check a 4-20 mA output or test analogue input/output cards.

24 V loop power supply

The 24 V loop power supply, up to 30 mA, makes it possible to power a two-wire transmitter directly from the communicator, even when it is disconnected from the plant, without external power supplies.

4-20 mA loop diagnostics

4-20 mA loop diagnostics make it possible to check the behaviour of the loop: abnormal current values, unstable signals or faulty wiring are among the most frequent causes of incorrect measurements. Carrying out the check with the same instrument that communicates with the transmitter speeds up fault finding.

Calibrating pressure transmitters in the field with the ADT161

The calibration of pressure transmitters in the field is one of the most frequent tasks for maintenance personnel. The ADT295 supports the optional external digital pressure module Additel ADT161, which covers from vacuum up to 60,000 psi (4,200 bar) and differential pressure up to ±1 inH₂O.

By combining the communicator with the module, you carry out a comparative calibration: the module measures the reference pressure, while the communicator, connected to the transmitter, reads its output and maintains HART communication throughout the procedure. You therefore work with a single portable instrument and avoid manual steps to compare values. For the pressure side, E Instruments also offers Additel pressure calibrators.

DD library and updates with no subscription

The DD library is the heart of a communicator: if an instrument's DD is missing, the communicator can only access the universal commands. The ADT295 HART communicator has over 2,600 pre-installed HART, FOUNDATION Fieldbus and Profibus PA devices and supports the latest FDI technology.

Library and firmware updates are free, with no subscription, and are carried out over Wi-Fi with a one-click procedure. This is an aspect worth considering carefully: with many communicators, updates are paid for and represent a recurring cost.

Field ruggedness and documentation management

A communicator often works outdoors, wearing gloves and in harsh environments. The ADT295 features:

  • a 5" colour touchscreen with physical keys, usable with gloves, with adjustable backlight and a tempered-glass protected display;
  • IP54 ingress protection and a 1-metre drop test;
  • operating temperature from -20 to 50 °C, 5 g vibration and 8 g shock;
  • battery life of about 20 hours in measurement-only use, with a high-capacity version available as an option.

For documentation, 64 GB of internal memory and built-in Bluetooth make it possible to keep manuals, wiring diagrams and work instructions on board and to share them directly in the field.

Additel ADT295: main features

  • Protocols: HART, FOUNDATION Fieldbus and Profibus PA, online and offline.
  • Display: 5" colour touchscreen with physical keys.
  • Built-in calibrator: current and voltage, accuracy 0.01% of reading.
  • Loop power supply: 24 V up to 30 mA, with internal resistor.
  • Pressure: optional ADT161 module, from vacuum to 4,200 bar.
  • DD library: over 2,600 devices, free updates over Wi-Fi.
  • Connectivity and memory: Wi-Fi, Bluetooth, 64 GB.
  • Ruggedness: IP54, -20 to 50 °C, 1 m drop.
  • Dimensions and weight: 192.3 × 96.5 × 55.3 mm, 0.8 kg.

How to choose a field communicator: the criteria to consider

Before choosing a HART and Fieldbus communicator, it is worth checking:

  1. The protocols present in your plant (HART only, or also Fieldbus and Profibus PA).
  2. The DD library and how it is updated (cost, frequency, FDI support).
  3. The measurement and generation functions you need (current, voltage, loop power supply).
  4. Pressure calibration, if required, and the availability of external modules.
  5. Environmental conditions: temperature, IP rating, use with gloves, display legibility.
  6. Documentation management in the field, to reduce intervention times.

Frequently asked questions about the HART and Fieldbus communicator

  1. What is the difference between a HART communicator and a multiprotocol communicator? The former communicates only with HART instruments; the latter also supports FOUNDATION Fieldbus and Profibus PA, and therefore covers more types of plant with a single instrument.
  2. Can a transmitter be calibrated without disconnecting it from the loop? Yes, in online mode the communicator communicates with the installed instrument; for pressure verification, an external reference module connected to the communicator is used.
  3. Is an external resistor needed to communicate in HART? Not with the ADT295, which includes the internal resistor and the loop power supply.

Discover the Additel ADT295 HART and Fieldbus communicator

To find out more about features and configurations, see the page for the Additel ADT295 field communicator and explore E Instruments' range of HART communicators.

Our technical team is available to analyse your application and help you choose the solution best suited to your maintenance needs.